Name: | 190~1100nm Double Beam UV Vis Spectrophotometer | Optical System: | Single, Littrow,Grating 1200 Line/mm |
---|---|---|---|
Wavelength Range: | 190~1100nm | Spectral Bandwidth: | 4nm |
Wavelength Accuracy: | ±0.5nm | Wavelength Repeatability: | ±0.2nm |
High Light: |
1100nm UV Vis Spectrophotometer,190nm UV Vis Spectrophotometer,1100nm double beam spectrophotometer |
190~1100nm Double Beam UV Vis Spectrophotometer with PC Analysis Software
Characteristics of UV Vis Spectrophotometer
Good stability: The integrated structure design (8mm heat-treated aluminum alloy base) ensures long-term reliability and stability of the instrument.
● High Accuracy: Using grating micrometer level precision screw drive to ensure the accuracy of wavelength <±0.5nm;Transmittance of accuracy reaches ±0.3%.
● Convenience for use: 5.7 inch graphic LCD display. Spectrum curve be clear at a glance and easy to operate. There is quantitative analysis,qualitative analysis, kinetics test, DNA/RNA, multi- wavelength analysis in special purpose testing program.
● Service life durable: Imported Halogen lamp ensures the light source for up two years of life, receiver life for 20 years.
●Wide application fields: Autosampler, micro cell holder,Optional Equipped with Peltier/sipper unit, with Micro cell holder, with (5°incident angle) Reflectance measurement attachment etc. expands the application areas of more.
Function of products
●Photometric measurement: Convenience to measure samples in the specified wavelength absorbance and transmittance. Multi-wavelength detector is maximum simultaneous determination of 10 wavelengths.
●Quantitative measurement: Establish automatically standard curve, first- order first after we order the third order curve fitting. Optional method: Single wavelength calibration,Double wavelength calibration,Three point method,
●Qualitative measurement: Including scaling, smoothing, filtering, detecting peak-valley etc.
●Kinetics measurement: Calculate kinetics of enzyme reaction rate. There are scaling, smoothing, filtering, detecting peak-valley and finding the derivative in a variety of graph processing methods for your choice.
●DNA/RNA,Test function of protein calculates automatically the sample ratio and concentration.
Technical Data of UV Vis Spectrophotometer
Model | NU-T6PC |
Optical System | Single, Littrow,Grating 1200 line/mm |
Wavelength Range | 190~1100nm |
Spectral Bandwidth | 4nm |
Wavelength Accuracy | ±0.5nm |
Wavelength Repeatability | ±0.2nm |
Photometric Accuracy | ±0.3%T;±0.002Abs(0~0.5Abs);±0.004(0.5-1.0Abs); |
Photometric Repeatability | ±0.1%T;±0.001Abs(0~0.5Abs);±0.002(0.5-1.0Abs); |
Stray Light | ≤0.05%T, @220nm/360nm |
Noise Level | ±0.002Abs |
Stability | ±0.002A/h@500nm |
Baseline Flatness | ±0.002Abs |
Baseline Dark Noise | 0.1%T |
Photometric Range | 0~200℅T,-4~4A,0~9999C(0-9999F) |
Light Source | Tungsten&Deuterium Lamp(Pre-aligned) |
Scan Speed | Hi,MED.,LOW.,MAX.3600nm/min |
Display | 320×240 Dots Array Graphic LCD |
System Language | English |
Output | USB1.1 Connecting to computer&Parallel port for printer |
Power Requirement | AC90~250V/ 50~60Hz |
Dimensions(L×W×H)(mm) | 550×400×210 |
Weight | 23kg |
Packing List of UV Vis Spectrophotometer
Description | Quantity |
Spectrophotometer | 1 |
Mains Lead | 1 |
Cuvettes | 4 Sets glass 2 Sets quartz |
Computer advanced application software | 1 |
Software manual | 1 |
Dongle | 1 |
Dust Cover | 1 |
Manual | 1 |